GATE Assistant

Past questions › Biomedical Engineering › GATE 2021

GATE 2021 · BM · NAT · 2 marks · Sensors and Bioinstrumentation › Medical Equipment Principles

GATE 2021 Biomedical Engineering question: A pacemaker was implanted in a cardiac patient. It has a battery of 2.4 A.h…

A pacemaker was implanted in a cardiac patient. It has a battery of 2.4 A.h (Ampere.hour). It is designed to deliver a rectangular pulse of amplitude 1.5 V for 1 ms ON time for every one second. The electrode – heart resistance is 150 . Assuming the current drained from the battery is negligible, the lifetime of the pacemaker is _____ years. (rounded off to the nearest integer)

Show answer

Official answer: 27 to 28