For the full bridge made of linear strain gages with gage factor 2 as shown in the diagram, at C and strain is 0. The temperature coefficient of resistance of the strain gages used is 0.005 per C. All strain gages are made of same material and exposed to same temperature. While measuring a strain of 0.01 at a temperature of 50 C, the output in millivolt is ______ (rounded off to two decimal places).

Show answer
Official answer: 2.45 to 2.55