GATE 2022 Biomedical Engineering question: A catheter based arterial blood pressure measurement device uses a flexible diaphragm…

A catheter based arterial blood pressure measurement device uses a flexible diaphragm mounted with four identical strain gauges in a Wheatstone bridge configuration as shown in the figure. Assume that the strain gauges have a nominal resistance value of = 10 kΩ, Gauge Factor G = 40 and Young’s Modulus E = 10 MPa. Blood pressure variations results in small finite change in strain ().

figure

If is the output voltage of the Wheatstone bridge and is the stress in MPa, the sensitivity is _______ V.MPa.

Show answer

Official answer: 8