GATE 2024 Biomedical Engineering question: A bio-potential signal of 4 mV on the skin surface was fed to an amplifier with a…

A bio-potential signal of 4 mV on the skin surface was fed to an amplifier with a differential gain of 2000. The noise in the signal is 1000 mV. If the amplifier output produces a noise output of 200 mV, the common mode rejection ratio of the amplifier is ______ dB. Give your answer rounded to the nearest integer.

Show answer

Official answer: 80